AED Recalls
| Alert Details for # 2006120002 (Released 12/01/2006 17:26 EST) |
| Description: |
Welch Allyn Portable Intensive Care (PIC) 50 System: Portable
monitor/defibrillator |
| Manufacturer: |
Medical Research Lab; Distributor: Welch Allyn Protocol, Inc., Beaverton, OR |
| Reason: |
Display of "ECG COMM" (ECG Communication) Error Message on display during use which may prevent or unacceptably delay delivery of defibrillation shock. |
| Analyst Comments: |
Please click on the attachment in the Links field to view notification details. Comply with Manufacturer instructions. Tech Support will have to service product. Pending service, product should undergo a Self-Test before each shift. Self-Test is automatically initiated by powering up the device. If the "ECG COMM" Error occurs during use, cycling the Power Off-On may clear the error. Even if power cycling clears the device, contact Manufacturer for priority servicing. Contact Welch Allyn Phone: 503-530-7500, Phone: 847-520-0300, Phone: 800-462-0777, Fax: 503-526-4200, or www.monitoring.welchallyn.com. |
| Alert Type: |
Recall |
Source Alert Type: |
Not Classified by Source |
| Domain: |
Biomedical Devices |
Source Type: |
US Army Medical Materiel Agency |
| Distribution: |
Not Specified |
Source Date: |
12/01/2006 |
| Product Description |
Model/Product Number |
Stock Identifiers |
Volume |
Recall ID |
| Welch Allyn Portable Intensive Care (PIC) 50 System: Portable monitor/defibrillator |
Part #s: 971081, 971082, 971083, 971084 |
Lot/Serial #: 80052, 80141, 80190, 80191, 80425, 80426, 80430, 80431, 80432, 80435, 80436, 80587, 80826, 80915, 80916, 80918, 80920, 80921, 80923 - 80930, 80932, 80934, 80960, 80962, 80963, 80967, 80969, 80971 - 80976, 80981, 80983, 80987, 80989, 80990, 80991, 80992, 80993, 81003 - 81019, 81020, 81022 - 81038, 81044 - 81047, 81049, 81070, 81094, 81102 - 81105, 81109 - 81113, 81165, 81167, 81178, 81179, 81181, 81184, 81187, 81188, 81192, 81194, 81197, 81258, 81297, 81499, 81627, 81765, 81767, 81769, 81770, 81773, 81803, 81804, 81823, 81827 - 81829, 81832 - 81835, 81837 - 81846, 81853, 81855, 81857, 81858, 81860, 81861 - 81869, 81884 - 81886, 81920 - 81946, 81949, 81957, 81983 - 81989, 81990, 81996 - 82005, 82016, 82017, 82019, 82021 - 82025, 82027, 82028, 82034 - 82039, 82043, 82061 - 82080, 82098 - 82109, 82111, 82122 - 82124, 82126 - 82128, 82160 - 82165, 82169, 82172 - 82186, 82192 - 82196, 82198 - 82202, 82204, 82252 - 82268, 82270 - 82273, 82284 - 82286, 82317 - 82319, 82321 - 82330, 82342 - 82344, 82346 - 82349, 82408, 82409, 82411 - 82419, 82421 - 82424, 82426, 82430, 82461 - 82475, 82517 - 82522, 82528 - 82533, 82536 - 82538, 82550, 82552, 82553, 82555 - 82557, 82559 - 82565, 82569, 82570, 82612 - 82619, 82621, 82623 - 82631, 82655 - 82672, 82689, 82690, 82773 - 82813, 82827 - 82830, 82852, 82857 - 82859, 82962, 82965 - 82967, 82975, 82986, 82989, 82991, 82998, 83010 - 83013, 83060 - 83067, 83084, 83088 - 83090, 83142, 83144, 83145, 83147. |
Not Specified |
MMQC-06-1370 |
|